![Agilent Technologies BioCel 1200 User Manual Download Page 21](http://html1.mh-extra.com/html/agilent-technologies/biocel-1200/biocel-1200_user-manual_2867701021.webp)
15
Chapter 2: BioCel overview
BioCel User Guide
BioCel specifications
About this topic
This topic describes the physical, electrical, and environmental
specifications of the BioCel.
Physical
specifications
Electrical
requirements for all
BioCels
Air-flow (pressure)
requirements
Equipment rating
The BioCel conforms to the requirements of Council Directive 72/23/
EEC relating to electrical safety by application of the following standard:
EN61010-1:1993, including Amendment 2:1995
Environmental
specifications
Dimension
1200 no hood
1600 with hood
1800 with hood
Height
196.8 cm (77.5 in)
259.1 cm (102 in)
259.1 cm (102 in)
Width
205.7 cm (81 in)
152.4 cm (60 in)
182.9 cm (72 in)
Depth
121.9 cm (48 in)
152.4 cm (60 in)
152.4 cm (60 in)
Weight
Approximately
1088.6 kg (2400 lb)
Approximately 2268 kg (5000 lb)
Requirement
System location
EU
US
Voltage
230V~
120V~
Frequency
50–60 Hz
50–60 Hz
Current
20 A
30 A
Fuse
20 A
30 A
Requirement
All BioCels
Air pressure
100–110 psi (0.62–0.69 MPa)
Requirement
1200
1600
1800
Temperature
15–40 °C
15–40 °C
15–40 °C
Relative
humidity
10–75% RH
10–75% RH
10–75% RH
Elevation
1–2000 m
1–2000 m
1–2000 m
Summary of Contents for BioCel 1200
Page 1: ...Agilent Technologies BioCel System User Guide...
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Page 18: ...Chapter 1 Introduction BioCel User Guide 12...
Page 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Page 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Page 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Page 104: ...Index BioCel User Guide 98...
Page 105: ......