Chapter 3: Preparing for a run
BioCel User Guide
Performing pre-run checks
About this topic
After turning on the BioCel, creating a protocol, and preparing the
devices you would typically perform a number of pre-run checks. This
topic describes the checks to perform.
Procedure
To check that the BioCel is ready for a run:
1.
For each VPrep, make sure that any reservoirs and Micro Wash trays
are on the correct shelves.
2.
Make sure that there are no stray plates or lids in robot-accessible
places, which include:
Platepads
VPrep shelves
VStack plate stages
VSpin buckets
VCode plate stages
PlateLoc plate stages
Plate hotel shelves
PlateHubs
Lid hotel bays
Lid removal devices
Third-party devices
You can also set the robot to check all accessible positions at the
beginning of a run.
!! DAMAGE HAZARD !!
Leaving a stray plate in the system
might cause the robot, when holding a plate, to crash into the
stray plate. A typical robot crash will damage the plates, spill
samples and stop the robot. It might also damage the robot
gripper pads.
Workflow this procedure belongs
to
“Workflow for preparing the BioCel for
a Run” on page 42
Preparing devices for a run
“Preparing devices and accessories” on
page 46
For information about...
See...
Summary of Contents for BioCel 1200
Page 1: ...Agilent Technologies BioCel System User Guide...
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Page 18: ...Chapter 1 Introduction BioCel User Guide 12...
Page 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Page 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Page 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Page 104: ...Index BioCel User Guide 98...
Page 105: ......