Chapter 3: Preparing for a run
BioCel User Guide
56
3.
For each VSpin, if you are using the robot to load the counterweight,
make sure that the counterweight plate is on the correct platepad in
the correct orientation and that the VSpin is empty.
If you are manually loading a counterweight, make sure that the
counterweight is loaded in the VSpin according to the protocol you
will run, and that it is oriented correctly.
4.
For each VStack, make sure that the stacker contains the plates
specified by the protocol and that the labware and plate name
displayed on the display panel match:
The actual labware in the stacker
The labware and plate name in the protocol
5.
Make sure that the device air pressures regulators lie within the
following ranges:
To find the current incoming air pressure for the PlatePierce, open
the PlatePierce Diagnostics software.
6.
For each VStack, open the VStack BioNet Diagnostics software to
check that the grip pressure is appropriate for the type of plates that
you are using.
For more information about adjusting the grip pressure, see the
VStack User Guide
.
7.
If your BioCel is equipped with environmental control, check that
environmental settings are within acceptable ranges for the run.
8.
Close all doors.
!! DAMAGE HAZARD !!
If running a BioCel 1600 or 1800,
make sure that you move the computer monitor out of the way
as you close the overhead doors.
9.
If the interlock override is on, put the key into the interlock override
keyhole in the power panel and turn it to the 12 o’clock position.
This disables the interlock override and stops the robot if the door is
opened.
Related topics
Device
Air Pressure (psi)
Air Pressure (MPa)
PlateLoc
87–90
0.60–0.62
VCode
85–90
0.56–0.62
VPrep
90–95
0.62–0.65
VStack
40–65
0.28–0.34
PlatePierce
75–80
0.65–0.55
For more information about...
See...
Workflow this procedure belongs
to
Summary of Contents for BioCel 1200
Page 1: ...Agilent Technologies BioCel System User Guide...
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Page 18: ...Chapter 1 Introduction BioCel User Guide 12...
Page 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Page 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Page 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Page 104: ...Index BioCel User Guide 98...
Page 105: ......