39
3 Installing BenchCel Microplate Handler
Verifying laboratory requirements
BenchCel Microplate Handler User Guide
Figure
Example of BenchCel on risers with six stacks showing height requirements of 250-mm, 660-mm, and
860-mm racks
Addition of devices
If you are integrating an Agilent device or third-party device with your BenchCel
Microplate Handler, make sure you include adequate space to accommodate these
devices. See the device user documentation for space requirement information.
Electrical requirements
The BenchCel Microplate Handler has the following power requirements. For power
requirements of other devices in an integrated workstation, see the applicable device
user documentation.
Compressed air requirements
The BenchCel Microplate Handler requires the use of clean, dry, compressed air to
move pneumatic components inside the device. The compressed air can be from the
following sources:
•
Centralized source (house)
•
Compressed-air cylinders
•
Portable pumps
250 mm
828 mm
1438 mm
1260 mm
660 mm
860 mm
Utility
Requirement
Electrical
100–240 V~, 50/60 Hz
Operating AC current
5 A at 120 V~, 2.5 A at 240 V~
Fuse
5 A, 250 V, 5
×
20 mm, fast acting