![Agilent Technologies BenchCel User Manual Download Page 154](http://html1.mh-extra.com/html/agilent-technologies/benchcel/benchcel_user-manual_2867698154.webp)
7 Maintenance and troubleshooting
Software error messages
144
BenchCel Microplate Handler User Guide
Timeout on grippers
The motor’s controller did not
respond as expected.
Home the robot, and then move the robot in
the same axis. If the robot moves, continue
on.
If the robot does not move, quit the VWorks
software and restart the BenchCel Microplate
Handler.
If the problem persists, contact Agilent
Technical Support.
Timeout on theta-axis
The motor’s controller did not
respond as expected.
Home the robot, and then move the robot in
the same axis. If the robot moves, continue
on.
If the robot does not move, quit the operating
software and restart the BenchCel Microplate
Handler.
If the problem persists, contact Agilent
Technical Support.
Timeout on x-axis
The motor’s controller did not
respond as expected.
Home the robot, and then move the robot in
the same axis. If the robot moves, continue
on.
If the robot does not move, quit the operating
software and restart the BenchCel Microplate
Handler.
If the problem persists, contact Agilent
Technical Support.
Timeout on z-axis
The motor’s controller did not
respond as expected.
Home the robot, and then move the robot in
the same axis. If the robot moves, continue
on.
If the robot does not move, quit the operating
software and restart the BenchCel Microplate
Handler.
If the problem persists, contact Agilent
Technical Support.
Error message
Cause
Solution