Verifying the installation and setup
Running the Shutdown utility
98
AssayMAP Bravo Platform Installation Guide
Running the Shutdown utility
You should run the Shutdown utility after every application protocol run, if the
AssayMAP Bravo Platform will not be in use for 1 hour or longer, and before
shutting down the device overnight.
The Shutdown utility prepares the AssayMAP Bravo Platform for idle time by:
–
Washing the syringes the specified number of wash cycles.
–
Aspirating 200 µL of Syringe Storage Liquid into the syringes. The
Syringe Storage Liquid prevents residual reagents from forming salts
and corroding the syringe seals.
Workflow
To run the Shutdown utility:
1
Open the
Utility Library
, locate the
System Startup/Shutdown
utility, and then
click
Utility
.
2
Click
Instructions
.
3
Follow the instructions to run the Shutdown utility.