Instructions for Use
Error Messages
16
MESSAGE DESCRIPTION
REPAIR
ACTION
In normal operation (after the starting
phase) the current drawn by the pump
is higher than programmed (8 A).
Check that the pump rotor is free
to rotate.
Press the START push-button
twice to start the pump.
After the starting phase the output
connection is shorted (output current
higher than 38 A).
Check connections and shortages
between pump and controller.
Press the START push-button
twice to start the pump.
The pump is stopped by an emergency
stop signal provided via a remote
contact.
Remove the controller power
cable and check the emergency
condition. Then reconnect the
power cable and
press the START push-button
twice
to start the pump.
The controller has received a random
signal or the line voltage is not
compatible with the setting of the
internal voltage switch.
Press the START push-button
twice to start the pump.
Should the message still be
present, call the Agilent service
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O
)
235/306