Technical Information
Use
17
N O R M A L O P E R A T I O N
X X K R P
M
where:
XX
= indicates the rotational speed (33 KRPM for high speed,
or 26 KRPM for low speed).
Operating the Pump
After the starting period, if the system has a vacuum leak or the
pressure in the pump chamber is high (from 1 mbar to atmosphere),
the pump continues to operate indefinitely. If the gas load at the
turbopump inlet flange continues to stay high, the power drawn by
the turbopump increases up to the maximum value.
Than the Turbo-V pump is slowed down in pro- portion to the gas
load at least until it reaches about 3 KRPM.
As soon as the gas load decreases, the pump will automatically
accelerate to reach normal operation. The pump can be stopped at
any rotational speed and can be restarted at any rotational speed
from either the front panel buttons or the remote connections. The
controller automatically synchronizes the output to the rotational
speed of the pump and then accelerates linearly up to the nominal
speed or within steps if the Soft Start has been selected.
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O)
293/306
Summary of Contents for 969-9462
Page 4: ...Turbo V 2000HT Controller 4 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 124: ...8 Bruksanvisning Felmeddelanden 124 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 152: ...10 Ohjekäsikirja Vianetsintä 152 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 208: ...14 Návod na Obsluhu Oznamy vád 208 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......