Technical Information
Controller Specifications
17
Controller Specifications
Tab. 3
Input:
Voltage
Frequency
Power
100, 120, 220, 240 Vac
±
10 %, 1-phase
50 - 60 Hz
1100 VA maximum
Output:
Voltage
Frequency
Start up power
Maximum output power:
Process gas Argon
Process gas Air
120 Vac nominal
±
10 %
550 Hz
±
2 %
700 W maximum
300 W
450 W
Operating temperature
0 °C to +40 °C
Storage temperature
-20 °C to +70 °C
Circuit breaker
10 A for 220 input Voltage
20 A for 120 input Voltage
P1 optically insulated input
Minimum ON 3mA Maximum 5mA
J2 optically insulated output
24 Vdc, 60 mA
J21 auxiliary power output
120 Vac, disregarding the mains
1 A for forepump relay coil, Vent Device (socket)
Radio interference
suppression
Conforms to EN 55011 class A group 1
Conforms to IEC1000-4-2,1000-4-3, 1000-4-4
Safety standard
Conforms to EN 61010-1
Auxiliary signals connectors
P1
J2
J13
External INPUT signals (pins)
OUTPUT signals (sockets)
RS 232/RS 422/RS 485 connection (optional)
Interconnecting cables
Mains cable (3-wire, 3-meter long)
Pump cable (6-wire, 5-meter long)
Weight (both models)
19 kg (42.56 lbs)
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O)
241/306
Summary of Contents for 969-9462
Page 4: ...Turbo V 2000HT Controller 4 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 124: ...8 Bruksanvisning Felmeddelanden 124 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 152: ...10 Ohjekäsikirja Vianetsintä 152 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 208: ...14 Návod na Obsluhu Oznamy vád 208 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......