17
Technical Information
Use
Where the sign minus (-) become a square (
) when the pump finish
the ramp-up step. The active step is indicated by a flashing
square (
).
As the ten steps are fully covered, the pump will reach the Normal
operation. If during the Soft Start mode the current drawn by the
pump exceed 600 W the speed of the pump is decreased to maintain
the maximum power allowable (600 W).
If the Soft Start mode has been deselected the display will change
and shows:
P U M P I S S T A R T I N G
1 2 X X K R P M
where:
1 2 = contrast inverted identifies the set point condition:
1 is displayed when relay K1 is de-energized and the related
output is zero voltage.
2 is displayed when relay K2 is energized and the related
output is 24V.
XX KRPM
= indicates the actual theoretical rotational speed of the
pump as a function of the controller output frequency (range 3 to 33
KRPM).
After START command, frequency output will be at the maximum
level, then the frequency will decrease to a value proportional to the
pump rotational speed (about 3 KRPM of nominal frequency if the
pump is completely stopped).
The pump will accelerate to its normal rotational speed.
During acceleration of the pump or during any operating condition,
it is always possible to select the other parameters to be displayed
pressing the PUMP CURRENT or the CYCLE NUMBER pushbuttons.
After the run up time and when the normal rotational speed is
reached, the display will be as follows, even if any previous display
selection was made, and the normal condition has been reached.
292/306
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O
)
Summary of Contents for 969-9462
Page 4: ...Turbo V 2000HT Controller 4 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 124: ...8 Bruksanvisning Felmeddelanden 124 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 152: ...10 Ohjekäsikirja Vianetsintä 152 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 208: ...14 Návod na Obsluhu Oznamy vád 208 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......