Technical Information
Description of the VacIon Pump
5
VacIon Plus 300 pumps User Manual
/
87-900-103-01 (H)
59/98
Description of the VacIon Pump
The Agilent VacIon Plus 300 pumps are ion pumps and are available
in three types:
StarCell
Noble Diode
Diode
They operate in the pressure range from 10
-2
to below 10
-11
mbar.
Virtually all gases and vapors can be pumped successfully with the
ion pump. The pumping speed will vary depending on the system
pressure, the gas type, and the applied operating voltage. The latter
can be optimized using Agilent controllers to achieve the lowest
possible operating pressure.
The VacIon Plus StarCell pump allows starting from as high as
5x10
-2
mbar, because ions are prevented from bombarding the system
and pump walls at starting pressures, and have high speed for inert
gases. The VacIon Plus Diode and Noble Diode pumps require a lower
starting pressure (i.e. 10
-4
mbar).
The tantalum in the Noble Diode pump enhances the noble gas
pumping capacity, although it somewhat reduces the high getterable
gas pumping speed that the standard diode pump provides.
The VacIon Plus StarCell is the latest variation of the Triode
configuration. Its patented design makes this ion pump the only one
that can handle a high amount of Noble Gases (better than Noble
Diode and Triode) and Hydrogen (comparable to the Diode).
A positive polarity, high voltage supply is required to operate Diode
and Noble Diode pumps. A negative polarity high voltage supply is
required to operate StarCell pumps since the anode is grounded and
the cathode is held at negative potential.
The inlet ConFlat flange is a 8” (NW 150).
Summary of Contents for 9191610
Page 3: ...VacIon Plus 300 VacIon Plus 300 pumps User Manual 87 900 103 01 H 3 98 VacIon Plus 300...
Page 4: ...VacIon Plus 300 4 98 VacIon Plus 300 pumps User Manual 87 900 103 01 H...
Page 8: ...Contents 8 98 VacIon Plus 300 pumps User Manual 87 900 103 01 H...
Page 56: ...4 Installation Procedure Disposal 56 98 VacIon Plus 300 pumps User Manual 87 900 103 01 H...
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......