142
Installation and User’s Guide
G
Network Analyzer Performance Specification Summary
Network Analyzer System Performance
The following specifications describe the system performance of the
Agilent E8364B PNA Series vector network analyzer Option 014 and UNL
configuration for measurements between 45 MHz and 50 GHz.
Calibration Kit
Agilent 85056A, 2.4 mm precision, with sliding loads
Cables
Agilent 85133F, 2.4 mm flexible test port cable set
Calibration Type
Full 2- port with sliding loads
Temperature
Range
25
°
C ±5
°
C
Warm-up Time
≥
0.5 hour
Maximum Output Power
Dynamic Range
System dynamic range is calculated as the difference between the noise
floor and the source maximum output power. Reflection measurements are
limited by directivity. Therefore, system dynamic range only applies to
transmission measurements.
Frequency Range (GHz)
0.045 to 10.0
10.0 to 20.0
20.0 to 30.0
30.0 to 40.0
40.0 to 45.0
45.0 to 50.0
Maximum output power
+3 dBm
0 dBm
−
4 dBm
−
8 dBm
−
11 dBm
−
17 dBm
Frequency Range (GHz)
0.045 to 0.5
0.5 to 2.0
2.0 to 10.0
10.0 to 20.0 20.0 to 30.0 30.0 to 40.0
40.0 to 45.0 45.0 to 50.0
System
dynamic
range
92 dB
117 dB
120 dB
119 dB
109 dB
105 dB
102 dB
95 dB
Summary of Contents for 85225F
Page 1: ...Agilent Technologies Agilent 85225F PerformanceModelingSystem Installation and User s Guide ...
Page 90: ...90 Installation and User s Guide 2 Installing the System ...
Page 102: ...102 Installation and User s Guide 3 Verifying System Functionality ...
Page 110: ...110 Installation and User s Guide 4 Servicing the System ...
Page 118: ...118 Installation and User s Guide A Enhancing Measurement Accuracy ...
Page 128: ...128 Installation and User s Guide C RF Subsystem Functional Verification Test ...