![Agilent Technologies 85225F Installation And User Manual Download Page 122](http://html.mh-extra.com/html/agilent-technologies/85225f/85225f_installation-and-user-manual_2867939122.webp)
122
Installation and User’s Guide
B
DC Subsystem Functional Verification Test
For Agilent E5260A/70B DC subsystems:
If you encounter a problem
✔
Refer to the Agilent E5260A/70B precision parametric measurement
solution user’s guide for an explanation of the error codes. See the
section titled “Error Codes” in the chapter titled, “If You Have a
Problem.”
✔
Refer to the troubleshooting information in the Agilent E5260A/70B
precision parametric measurement solution service manual.
To perform the Agilent E5260A/70B precision parametric measurement solution self-test
Step
Notes
1 Disconnect all cables from the
measurement terminals on the
E5260A/70B front panel.
2 Connect the E5260A/70B to line
power.
3 Switch on the instrument line
power switch.
4 Wait 1 hour before continuing to
step 5
.
5 Press Shift > Menu.
6 Move the cursor to SELFTEST and
press Enter. Move the cursor to
EXECUTE and press Enter.
7 Use the arrow keys to select ALL
and press Enter. At the completion
of the test, press Exit three times to
exit the setup menu.
8 To display the results, move the
cursor to SELFTEST and press
Enter. Move the cursor to RESULT
and press Enter.
9 Use the arrow keys to select
FRAME or Slot n.
The test result is displayed for each
item.
Summary of Contents for 85225F
Page 1: ...Agilent Technologies Agilent 85225F PerformanceModelingSystem Installation and User s Guide ...
Page 90: ...90 Installation and User s Guide 2 Installing the System ...
Page 102: ...102 Installation and User s Guide 3 Verifying System Functionality ...
Page 110: ...110 Installation and User s Guide 4 Servicing the System ...
Page 118: ...118 Installation and User s Guide A Enhancing Measurement Accuracy ...
Page 128: ...128 Installation and User s Guide C RF Subsystem Functional Verification Test ...