A-4
85052C
Standard Definitions
Standard Class Assignments
IMPORTANT
The following calibrations are only supported by certain PNA analyzers. See
your PNA network analyzer embedded help system.
Notes:
1. If you are performing a TRL calibration:
• S
21
T and S
12
T must be defined as thru standards.
• S
11
A and S
22
A must be defined as reflection standards.
• S
11
B, S
11
C, S
22
B, and S
22
C must be defined as line standards.
2. If you are performing a TRM calibration:
• S
21
T and S
12
T must be defined as thru standards.
• S
11
A and S
22
A must be defined as reflection standards.
• S
11
B, S
11
C, S
22
B, and S
22
C must be defined as match standards.
3. If you are performing an LRM calibration:
• S
21
T and S
12
T must be defined as line standards.
• S
11
A and S
22
A must be defined as reflection standards.
• S
11
B, S
11
C, S
22
B, and S
22
C must be defined as match standards.
4. S
11
B and S
11
C must be defined as the same standard.
5. S
22
B and S
22
C must be defined as the same standard.
Table A-3
Standard Class Assignments for the PNA Series Network Analyzer
Calibration Kit Label:
3.5 mm TRL Model 85052C
Class
A
a
S
11
A
1
S
11
B
2
S
11
C
2
S
21
T
4
S
22
A
1
S
22
B
2
S
22
C
2
S
12
T
4
a. For additional ports, make sure values
match the correct sex of the port.
Summary of Contents for 85052C
Page 5: ...1 1 1 General Information...
Page 13: ...2 1 2 Specifications...
Page 19: ...3 1 3 Use Maintenance and Care of the Devices...
Page 33: ...4 1 4 Performance Verification...
Page 36: ...4 4 85052C Performance Verification Recertification...
Page 37: ...5 1 5 Troubleshooting...
Page 41: ...6 1 6 Replaceable Parts...
Page 45: ...A 1 A Standard Definitions...
Page 62: ...Index 4 85052C Index part number 6 3 Z zeroing connector gage 3 8 3 10...