Standard Definitions
Standard Class Assignments
Standard Class Assignments
Class assignment organizes calibration standards into a format compatible with the error
models used in the measurement calibration. A class or group of classes corresponds to the
systematic errors to be removed from the measured network analyzer response.
through
list the classes of the devices in the kit for various network analyzers. This
information resides on the calibration definitions disk included in the kit.
Table A-1
Standard Class Assignments for the 8510 Network Analyzer
Disk File Name: CK_35MMC4
Calibration Kit Label: 3.5 mm C.4
Class
A
B
C
D
E
F
G
Standard Class Label
S
11
A
2
Open
S
11
B
1
Short
S
11
C
9
Load
S
22
A
2
Open
S
22
B
1
Short
S
22
C
9
Load
Forward transmission
11
Thru
Reverse transmission
11
Thru
Forward match
11
Thru
Reverse match
11
Thru
Forward isolation
a
a. The forward isolation standard is also used for the isolation part of the response and isolation
calibration.
9
Isol’n Std
Reverse isolation
9
Isol’n Std
Frequency response
1
2
11
Response
TRL thru
11
Thru
TRL reflect
1
Short
TRL line
15
16
17
Lines
Adapter
13
Adapter
TRL Option
Cal Z
0
:
_____ System Z
0
__X__ Line Z
0
Set ref:
__X__ Thru
_____ Reflect
Lowband
frequency: __2.0
GHz__
Summary of Contents for 85052C
Page 5: ...1 1 1 General Information...
Page 13: ...2 1 2 Specifications...
Page 19: ...3 1 3 Use Maintenance and Care of the Devices...
Page 33: ...4 1 4 Performance Verification...
Page 36: ...4 4 85052C Performance Verification Recertification...
Page 37: ...5 1 5 Troubleshooting...
Page 41: ...6 1 6 Replaceable Parts...
Page 45: ...A 1 A Standard Definitions...
Page 62: ...Index 4 85052C Index part number 6 3 Z zeroing connector gage 3 8 3 10...