3-6
85052C
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Connectors
The gages available from Agilent Technologies are intended for preventive maintenance
and troubleshooting purposes only. They are effective in detecting excessive center
conductor protrusion or recession, and conductor damage on DUTs, test accessories, and
the calibration kit devices.
Do not use the gages for precise pin depth measurements. See
for part number information.
Connector Gage Accuracy
The connector gages are only capable of performing coarse measurements. They do not
provide the degree of accuracy necessary to precisely measure the pin depth of the kit
devices. This is partially due to the repeatability uncertainties that are associated with the
measurement. Only the factory—through special gaging processes and electrical testing—
can accurately verify the mechanical characteristics of the devices.
With proper technique, the gages are useful in detecting gross pin depth errors on device
connectors. To achieve maximum accuracy, random errors must be reduced by taking the
average of at least three measurements having different gage orientations on the
connector. Even the resultant average can be in error by as much as
±
0.0001 inch due to
systematic (biasing) errors usually resulting from worn gages and gage masters. The
information in
assumes new gages and gage masters. Therefore,
these systematic errors were not included in the uncertainty analysis. As the gages
undergo more use, the systematic errors can become more significant in the accuracy of the
measurement.
The measurement uncertainties in are primarily a function of the assembly materials and
design, and the unique interaction each device type has with the gage. Therefore, these
uncertainties can vary among the different devices. For example, note the difference
between the uncertainties of the opens and shorts in
.
The observed pin depth limits in
add these uncertainties to the typical factory
pin depth values to provide practical limits that can be referenced when using the gages.
See
. Refer to
for more information on
the design of the calibration devices in the kit.
NOTE
When measuring pin depth, the measured value (resultant average of three
or more measurements) is not the true value. Always compare the measured
value with the observed pin depth limits in
to evaluate the
condition of device connectors.
Summary of Contents for 85052C
Page 5: ...1 1 1 General Information...
Page 13: ...2 1 2 Specifications...
Page 19: ...3 1 3 Use Maintenance and Care of the Devices...
Page 33: ...4 1 4 Performance Verification...
Page 36: ...4 4 85052C Performance Verification Recertification...
Page 37: ...5 1 5 Troubleshooting...
Page 41: ...6 1 6 Replaceable Parts...
Page 45: ...A 1 A Standard Definitions...
Page 62: ...Index 4 85052C Index part number 6 3 Z zeroing connector gage 3 8 3 10...