3-12
85052C
Use, Maintenance, and Care of the Devices
Connections
Connections
Good connections require a skilled operator. The most common cause of measurement error
is bad connections. The following procedures illustrate how to make good connections.
How to Make a Connection
Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded,
conductive table mat. Refer to
“Electrostatic Discharge” on page 3-2
for ESD
precautions.
2. Visually inspect the connectors. Refer to
“Visual Inspection” on page 3-3
.
3. If necessary, clean the connectors. Refer to
“Cleaning Connectors” on page 3-4
.
4. Use a connector gage to verify that all center conductors are within the observed pin
. Refer to
“Gaging Connectors” on page 3-6
5. Carefully align the connectors. The male connector center pin must slip concentrically
into the contact finger of the female connector.
6. Push the connectors straight together and tighten the connector nut finger tight.
CAUTION
Do not turn the device body. Only turn the connector nut. Damage to the
center conductor can occur if the device body is twisted.
Do not twist or screw the connectors together. As the center conductors mate, there is
usually a slight resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make
uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both
mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the
connection from long or heavy devices or cables.
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection.
provides information about the
torque wrench recommended for use with the calibration kit. A torque wrench is included
in the calibration kit. Refer to
for replacement part number and
ordering information.
Table 3-1
Torque Wrench Information
Connector Type
Torque Setting
Torque Tolerance
3.5 mm
90 N-cm (8 in-lb)
±
9.0 N-cm (
±
0.8 in-lb)
Summary of Contents for 85052C
Page 5: ...1 1 1 General Information...
Page 13: ...2 1 2 Specifications...
Page 19: ...3 1 3 Use Maintenance and Care of the Devices...
Page 33: ...4 1 4 Performance Verification...
Page 36: ...4 4 85052C Performance Verification Recertification...
Page 37: ...5 1 5 Troubleshooting...
Page 41: ...6 1 6 Replaceable Parts...
Page 45: ...A 1 A Standard Definitions...
Page 62: ...Index 4 85052C Index part number 6 3 Z zeroing connector gage 3 8 3 10...