275
Error Messages
Display Messages
F.1 Display Messages
FAIL
nnnn
indicates that the self test has failed. The number nnnn is a four
digit hexadecimal number that indicates which part of the self test
has failed.
So
FAIL0010
would mean that the DSP (Digital Signal Processor)
Communications had failed,
FAIL0012
would mean that the DSP
Communications had failed, and so had the Battery RAM.
Bits Mnemonics
Hexadecimal
Value
8
Counter
0100
16
7
Analog to Digital Convertor
0080
16
6
General DSP Hardware
0040
16
5
DSP Timeout
0020
16
4
DSP Communications
0010
16
3
Calibration Data
0008
16
2
Keypad
0004
16
1
Battery RAM
0002
16
0
Calibration Data Checksum
0001
16