170
Specifications
Specifications
λ
:
Entering of wavelength for automatic correction of attenuation
using typical correction values.
Cal:
Offset factor to adjust the attenuation factor on the display
within
±
99.999dB range.
Disp
→
Cal:
Sets attenuation value on the display to 0.000dB.
Swp:
Manual or automatic up or down attenuation sweep. Start,
stop, step size and dwell time (not for manual sweep) can be
entered.
Back Refl:
Desired return loss (back reflection level) can be
entered. Requires Agilent 81000BR back reflector, or Option 203.
Enb/Dis:
Optical signal path interrupted with shutter (>80dB
isolation).
Store/Recall:
9 user-selectable parameter settings may be stored
and recalled. Recall of default setting.
General
Recalibration period: 1 year.
Warm-up time: 45 Minutes. Not required if previously stored
within operating temperature range.
GPIB Capability:
All modes and parameters can be
programmed, SCPI command set, 8157A compatibility mode.
GPIB Interface Function Code:
SH1, AH1, T6, L4, SR1,
RL1, PP0, DC2, DT0, C0
Environmental
Storage temperature: -40 to +70
°
C
Operating temperature: 0 to +55
°
C
Humidity: <95% R.H. (to 40
°
C)
Altitude: to 10,000 feet