112
Remote Commands
OUTPut Commands
Example
OUTPUT 728;":INP:ATT?"
ENTER 728; Att
OUTPUT 728;":INP:OFFS?"
ENTER 728; Cal
Newcal = Basepow - Att + Cal
OUTPUT 728;":INP:OFFS ";Newcal
OUTPUT 728;":OUTP:APM ON"
OUTPUT 728;":OUTP:APM?"
ENTER 728;A$
:OUTPut:POWer
Syntax
:OUTPut:POWer
<wsp>
<value>[
DBM
]|
MIN
|
DEF
|
MAX
Description
This command sets the through-power for the
instrument. The through-power is used to set
the attenuation of the filter.
Att
filter
(dB) = ThroughPower
Base
(dBm) - ThroughPower(dBm) + Att
filter@Base
(dB)
You set the through-power by sending a value
(default units are dBm), or by sending
MIN
,
DEF
or
MAX
, which specify the minimum,
default and maximum values for the through-
power.
The maximum value and the default value are
those values for which Attenuation
filter
= 0dB.
The minimum value is that value for which
Attenuation
filter
is at its greatest. For example,
if you have set
INP:ATT 10
and
INP:OFFS
2
and then switched
UTP:APM ON
, then the
through power is set to 12dBm. The maximum
through power, and the default through power,
in this case is 22dBm. The minimum through
power in this case is -38dBm.