![Agilent Technologies 8156A Operating And Programming Manual Download Page 107](http://html.mh-extra.com/html/agilent-technologies/8156a/8156a_operating-and-programming-manual_2868603107.webp)
106
Remote Commands
INPut Commands
8.5 INPut Commands
:INPut:ATTenuation
Syntax
:INPut:ATTenuation
<wsp>
<value>[
DB
]|
MIN
|
DEF
|
MAX
Description
This command sets the attenuation factor for
the instrument. The attenuation factor is used,
with the calibration factor (see ) to set the filter
attenuation.
Attenuation
filter
(dB) = Att(dB) - Cal(dB)
You set the attenuation factor by sending a
value (default units are dB), or by sending
MIN
,
DEF
or
MAX
, which specify the
minimum, default and maximum values for the
attenuation factor.
The minimum value and the default value are
those values for which Attenuation
filter
= 0dB.
The maximum value is that value for which
Attenuation
filter
is at its greatest.
:INPut:ATTenuation?
Syntax
:INPut:ATTenuation?
[<wsp>
MIN
|
DEF
|
MAX
]
Description
The query returns the current attenuation
factor, in dB.
Attenuation
filter
(dB) = Att(dB) - Cal(dB)
By sending
MIN
,
DEF
, or
MAX
with the query
the minimum, default or maximum value
possible for the attenuation factor is returned.
Example
OUTPUT 728;":INP:ATT 32.15"