Remote Commands
OUTPut Commands
calibration factor (see ) to set the attenuation
factor to the required value for use as the base
value for the through-power
Cal
New
= (Through - Power
Base
- Att) + Cal
Current
When you switch the absolute power mode
OFF
, the last set calibration factor becomes
active, and the attenuation factor is set so that
the filter attenuation does not change. That is
Att(dB) = Attenuation
filter
(dB) + Cal(dB)
N O T E
Using any of the
:INPut:ATTenuation
commands or queries, or
any of the
:INPut:OFFSet
commands or queries, switches the
absolute power mode off automatically.
See “:OUTPut:POWer” on page 112 for
information on setting the through-power.
Switch the mode off (using
OFF
or
0
) to set the
attenuation of the filter by specifying the
attenuation and calibration factors. Switch the
mode on (using
ON
or
1
) to set the attenuation
of the filter by specifying the through-power.
:OUTPut:APMode?
Syntax
:OUTPut:APMode?
Description
The query returns whether the attenuation of
the filter is set by the attenuation and
calibration factors, or by the through-power.
0
indicates the instrument sets the attenuation
of the filter from the attenuation and calibration
factors.
1
indicates that the instrument sets the
attenuation of the filter from the through-
power.