![Agilent Technologies 8156A Operating And Programming Manual Download Page 120](http://html.mh-extra.com/html/agilent-technologies/8156a/8156a_operating-and-programming-manual_2868603120.webp)
119
Remote Commands
STATus Commands
Description
This command sets the bits in the PTRansition
register. Setting a bit in this register enables a
positive transition (0
→
1) in the corresponding
bit in the CONDition register to set the bit in
the EVENt register.
:STATus:OPERation:PTRansition?
Syntax
:STATus:OPERation:PTRansition?
Description
This query returns the current contents of the
OPERation:PTRansition register.
Example
OUTPUT 728;":STAT:OPER:PTR 138"
OUTPUT 728;":STAT:OPER:PTR?"
ENTER 728;A$
:STATus:QUEStionable:CONDition?
Syntax
:STATus:QUEStionable:CONDition?
Description
This query reads the contents of the
QUEStionable:CONDition register. Only one
bit of the condition register is used:
•
Bit 8, which is 1 when the wavelength is not
within the range of the user wavelength
calibration data.
Example
OUTPUT 728;":STAT:QUES:COND?"
ENTER 728;A$
:STATus:QUEStionable:ENABle
Syntax
:STATus:QUEStionable:ENABle
<wsp> <value>
Description
This command sets the bits in the ENABle
register that enable the contents of the EVENt
register to affect the Status Byte (STB). Setting
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com