5
General Maintenance
5975/5977 Series MSD Operating Manual
139
To Remove the EI XTR, SS, or Inert Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
3
Disconnect the wires from the EI source. Do not bend the wires any more
than necessary. (See
“To Connect/Disconnect Wiring from the EI XTR, SS,
4
Remove the thumbscrews that hold the ion source in place. (See
5
Pull the ion source out of the source radiator.
WARNING
The analyzers, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber
CAUTION
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not the wires.