5
General Maintenance
144
5975/5977 Series MSD Operating Manual
To Disassemble the EI XTR Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
Refer to the exploded parts view
and the EI EXT parts list
page 145 while using this procedure.
1
Remove the EI XTR source. (See
“To Remove the EI XTR, SS, or Inert
2
Remove the filaments by removing the two gold plated screws and
separating the filaments from the source.
3
Loosen the two gold plated screws on the source heater block assembly, and
separate the repeller assembly from the source body. The repeller assembly
includes the source heater block assembly, repeller, and related parts.
4
Remove the gold plated setscrews from the side of the source body.
5
Pull the entrance lens and ion focus lens to remove them from the source
body.
6
Remove the extractor lens and insulator.
7
Separate the entrance lens and ion focus lens from the lens insulator.
8
Remove the repeller nut, washers, and insulator from the front side of the
source heater block assembly, then remove the repeller, insulator, and
repeller block insert from the opposite side.