Operating in EI Mode
5975/5977 Series MSD Operating Manual
79
To view the firmware version of the MSD from
the GC control panel
1
Press
MS/Aux Det
to bring up the 5977B MSD menu.
2
Press the down arrow to scroll to
Firmware
.
To view the serial number of the MSD from the
GC control panel
1
Press
MS/Aux Det
to bring up the 5977B MSD menu.
2
Press the down arrow to scroll to
Serial#
.
To configure the network settings for the MSD
from the GC control panel
1
Press
Config,
and then press
MS/Aux Det
to bring up the CONFIGURE MS
DETECTOR menu.
2
To configure the
IP:
parameter, use the GC keypad to enter the new IP
address for the MSD, then press
Enter
to complete the entry.
3
Wait for the GC to display the new IP address. Reboot the MSD or proceed to
the gateway address with the down arrow button
4
Press the down arrow to scroll to
GW:
and use the GC keypad to enter the
new gateway address for the LAN and press
Enter
to complete the entry.
5
Press the down arrow to scroll to
SW:
and use the GC keypad to enter the
new subnet mask for the LAN and press
Enter
to complete the entry.
6
Reboot the MSD. (See below)
To reboot the MSD from the GC control panel
1
Press
Config,
and then press
MS/Aux Det
to bring up the CONFIGURE MS
DETECTOR menu.
2
Press the down arrow to scroll to
Request MSD Reboot?
.
3
Press
On/Yes
to reboot the MSD and wait for the MSD to complete this cycle
before trying to access it.