5
General Maintenance
5975/5977 Series MSD Operating Manual
117
Maintaining the Analyzer
Scheduling
None of the analyzer components require periodic maintenance. Some tasks,
however, must be performed when MSD behavior indicates they are necessary.
These tasks include:
•
Cleaning the ion source
•
Replacing filaments
•
Replacing the electron multiplier horn
The
Agilent 5977B Series MSD Troubleshooting and Maintenance Manual
provides information about symptoms that indicate the need for analyzer
maintenance. The troubleshooting material in the online help in the Agilent
MassHunter GC/MS Acquisition software provides more extensive information.
Precautions
Cleanliness
Keep components clean during analyzer maintenance. Analyzer maintenance
involves opening the analyzer chamber and removing parts from the analyzer.
During analyzer maintenance procedures, take care to avoid contaminating the
analyzer or the interior of the analyzer chamber. Wear clean gloves during all
analyzer maintenance procedures. After cleaning, parts must be thoroughly
baked out before they are reinstalled. After cleaning, analyzer parts should be
placed only on clean, lint-free cloths.
CAUTION
If not done correctly, analyzer maintenance can introduce contaminants into
the MSD.
WARNING
The analyzer operates at high temperatures. Do not touch any part until you
are sure it is cool.