Installing 7890/7820/6890/6850 GC Columns
40
5975/5977 Series MSD Operating Manual
15
Gently
check the alignment of the ion source and the interface tip seal.
When the ion source is aligned correctly, the analyzer chamber can be
closed all the way with no resistance except the spring tension from the
interface tip seal.
16
You can align the ion source and interface tip seal by wiggling the side plate
on its hinge. If the door still will not close, contact your Agilent Technologies
service representative.
17
Close the analyzer chamber door. (See
“To Close the Analyzer Chamber”
CAUTION
Forcing the analyzer door closed if these parts are misaligned will damage the
tip seal or the interface or the ion source, or will keep the sideplate from
sealing.