Intuvo 9000 GC Column Maintenance
68
5975/5977 Series MSD Operating Manual
To Condition an Intuvo Capillary Column
Materials needed
•
Carrier gas, (99.9995% pure or better)
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Tweezers (8710-2460)
•
Intuvo torque driver (5190-9571)
•
Intuvo gasket, polyimide 5/pk, for temperatures < 350 °C (5190-9072)
•
Intuvo gasket, nickel 5/pk, for temperatures 350 °C to 450 °C (5190-9073)
•
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Procedure
1
Install the column that requires conditioning. (See
2
Set a minimum velocity of 30 cm/s, or as recommended by the column
manufacturer. Allow gas to flow through the column at room temperature for
15 to 30 minutes to remove air.
3
Increase the column temperature to 120 °C.
4
Hold at the this temperature for 30 minutes.
5
Use MassHunter to run an air and water check. Proceed to the next step if the
air and water are within limits.
WARNING
The inlet, detector, bus components, and column can be hot enough to cause
burns. Cool heated zones to a safe handling temperature before continuing.
CAUTION
Always wear clean gloves while handling any parts that are in the flow path
going to the analyzer or parts that are located inside the analyzer.