
5
General Maintenance
140
5975/5977 Series MSD Operating Manual
To Connect/Disconnect Wiring from the EI XTR,
SS, and Inert Sources
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Pliers, long-nose (8710-1094)
•
Tweezers (8710-2460)
Procedure
1
Use tweezers or needle nose pliers to connect/disconnect the ceramic
source board wire leads at the source connectors. (See
.)
2
Use tweezers or needle nose pliers to connect/disconnect the source heater
wire leads at the ceramic source board.
Figure 43.
Wiring for the EI XTR Source
Extractor lens
Source heater
Repeller
Filament 1
Source RTD
Filament 2
Ceramic source board
Entrance lens
Ion focus lens