Contacting
Electrode
Method
(used
with
Rigid
Metal
Electrode
)
This
method
uses
Rigid
electrodes
which
make
contact
directly
the
surface
of
the
test
material.
This
method
is
applicable
for
thick,
smooth
or
slightly
compressible
materials
.
The
merits
and
demerits
of
this
method
are
as
follows:
Merits
Procedure
to
measure
capacitance
is
simple
It
is
not
necessary
to
apply
thin
lm
electrodes
Equations
to
obtain
dielectric
constant
are
simple
Demerits
Air
lm
(error
caused
by
air
gap
between
electrodes
and
surface
of
the
test
material)
causes
error
.
Principle
Figure
3-7
shows
the
schematic
electrode
structure
for
this
method.
Figure
3-7.
Contacting
Electrode
Method
(Rigid
Metal
Electrode)
Operation
3-11