Chapter 5
61
Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
5. S
pec
ific
ati
ons
an
d
S
upp
lem
ent
a
l
P
er
fo
rm
anc
e C
har
ac
ter
isti
cs
Supplemental Performance Characteristics
This section provides useful data on the 16196A/B/C/D. These supplemental performance
characteristics should not be considered specifications.
Additional Error
Additional errors are calculated as follows.
|Z| Measurement
Additional error for Impedance Ze [%] is calculated by substituting the values in the table
below into the following equation.
Ze
[%] = ± {A + (Zs/Zx + Yo ×Zx) × 100}
where
A [%]
Test Fixture’s Proportional Error [%]
Yo [S]
Test Fixture’s Open Repeatability [S]
Zs [
Ω
]
Test Fixture’s Short Repeatability [
Ω
]
Zx [
Ω
]
Measured Impedance Value of DUT [
Ω
]
where f is frequency (GHz).
D Measurement
Additional error for Dissipation Factor De is calculated by using the additional error for
Impedance Ze [%] as follows.
If Dx
≤
0.1:
De = Ze / 100
If 0.1
<
Dx
≤
0.5:
De = (Ze / 100) × (1 + Dx)
where Dx is the measured value of D. It is necessary for Ze to be below 10 %.
NOTE
D is not expressed as a percentage but as an absolute value.
Zs
(30 + 125 × f ) × 10
-3
[
Ω
]
Yo
(5 + 40 × f ) × 10
-6
[S]
A
1 × f
2
[%]
Summary of Contents for 16196A
Page 7: ...6...
Page 18: ...17 2 Product Overview 2 Product Overview...
Page 37: ...36 Chapter3 Operation Connecting and Measuring DUTs...
Page 38: ...37 4 User Maintenance 4 User Maintenance...
Page 41: ...40 Chapter4 User Maintenance Overview Figure 4 1 Flowchart of User Maintenance...
Page 59: ...58 Chapter4 User Maintenance Assembling Check...