![Agilent Technologies 16196A Operation And Service Manual Download Page 27](http://html1.mh-extra.com/html/agilent-technologies/16196a/16196a_operation-and-service-manual_2867577027.webp)
26
Chapter 3
Operation
Selecting and Changing the Insulator Assembly
NOTE
The number of the insulator assembly doesn’t indicate the maximum diameter of the DUT
insertion hole that can insert the cylindrical device.
NOTE
If the gap between the DUT and the insulator is large, the measurement accuracy and
repeatability decrease. Select an insulator assembly that is appropriate for the shape of the
DUT to be measured.
16196B
φ
0.85
1.0
0.5
0.5
φ
0.75
1.0
0.5
0.35
φ
0.68
1.0
0.5
0.35
16196C
φ
0.48
0.6
0.3
0.3
16196D
φ
0.30
0.4
0.2
0.13 or
0.2
*1
φ
0.34
0.4
0.2
0.2
*1.When you measure a 0.4 x 0.2 x 0.2 DUT, first check whether it
fits in the
φ
0.30 insulator assembly. Only if it does not, use the
φ
0.34 insulator assembly.
Table 3-1
Insulator Assembly Specifications
Insulator
Assembly
Example of Corresponding Chip (mm)
Length (L)
Width (W)
Height (H)