Chapter 4
49
User Maintenance
Check Sheet
4.
Us
er
Mai
nte
nan
ce
Check Sheet
Check Sheet Fill-Out Example
The following example shows how the check sheet is filled out following electrode wear
check. Fill out the sheet in the same manner for short plate wear check.
Electrode Wear Check Fill-Out Example
Table 4-3
Reference Value and User Limit Values Fill-Out Example
Frequency
*1
Measurement
Parameter
Reference
Value
*2
[a]
User Limit
Value
*3
[b]
Lower Limit
[a-b]
Upper Limit
[a+b]
100 MHz
Rs
90
m
Ω
30
m
Ω
60
m
Ω
120
m
Ω
Ls
−
290
pH
500
pH
−
790
pH
210
pH
800 MHz
Rs
310
m
Ω
40
m
Ω
270
m
Ω
350
m
Ω
Ls
−
260
pH
400
pH
−
660
pH
140
pH
*1.Set by the user as desired.
*2.Record values obtained at the time of “Reference Value Acquisition” on page 46.
*3.See also “Example of User Limit Values Setting” on page 44.
Table 4-4
Check History Fill-Out Example
Date
Frequency
Measurement
parameter
Measured Value
Pass/Fail
Oct./11/1999
9:30
100 MHz
Rs
100
m
Ω
Pass
Ls
−
320
pH
Pass
Oct./11/1999
9:35
800 MHz
Rs
345
m
Ω
Pass
Ls
−
360
pH
Pass
Oct./12/1999
9:30
100 MHz
Rs
105
m
Ω
Pass
Ls
−
340
pH
Pass
Oct./12/1999
9:35
800 MHz
Rs
355
m
Ω
Fail
*1
Ls
−
320
pH
Pass
*1.When the result is unacceptable, replace the part.
Summary of Contents for 16196A
Page 7: ...6...
Page 18: ...17 2 Product Overview 2 Product Overview...
Page 37: ...36 Chapter3 Operation Connecting and Measuring DUTs...
Page 38: ...37 4 User Maintenance 4 User Maintenance...
Page 41: ...40 Chapter4 User Maintenance Overview Figure 4 1 Flowchart of User Maintenance...
Page 59: ...58 Chapter4 User Maintenance Assembling Check...