56
Chapter 4
User Maintenance
Assembling Check
Short Plate Check
Measuring the fixture’s impedance with the short plate in place should check the condition
of the short plate.
CAUTION
“Short Plate Check” should be performed after the “Electrode Check” has been completed.
Required Tools
•
1.5-mm hex wrench (provided accessory)
•
Impedance measuring instrument (with 7-mm connector and calibrated)
NOTE
To make the same settings as given in the above table, refer to the Operation Manual for
the measuring instrument.
Procedure
Step 1.
Clean the short plate as described in “Cleaning” on page 41.
Step 2.
Remove the cap and place the short plate with the protruding surface down on the insulator
assembly.
Step 3.
Place the cap on the fixture body and fasten it.
Step 4.
Measure Rs and Ls at 100 MHz and 1 GHz in this condition.
Step 5.
Confirm that the Rs and Ls values are within the representative values given in the table
below. If the results are outside the limit range, please replace the short plate.
Table 4-14
Setting of Measuring Instrument
Measurement
condition
Set Value
Electrical Length
16196A: 26.2 mm
16196B: 26.9 mm
16196C: 27.1 mm
16196D: 27.3 mm
Measurement Parameter
Ls, Rs
OSC Level
500 mV
Point Averaging
32
Summary of Contents for 16196A
Page 7: ...6...
Page 18: ...17 2 Product Overview 2 Product Overview...
Page 37: ...36 Chapter3 Operation Connecting and Measuring DUTs...
Page 38: ...37 4 User Maintenance 4 User Maintenance...
Page 41: ...40 Chapter4 User Maintenance Overview Figure 4 1 Flowchart of User Maintenance...
Page 59: ...58 Chapter4 User Maintenance Assembling Check...