Chapter 3
Operation
Flow of Measurements
Flow of Measurements
Follow the steps below when performing measurements of DUTs with the 16196A/B/C/D.
1. Selecting and Changing the Insulator Assembly
Select an insulator assembly that is appropriate for the shape of the measured DUT and
replace the insulator assembly in the fixture.
2. Setting the Electrical Length
Set the fixture's electrical length in the measuring instrument you will be using.
3. Perfoming Fixture Compensation
Measure the data for open compensation and measure the data for short compensation.
When performing measurements with higher precision, carry out "Fixture
compensation for higher precision measurements".
4. Connecting and Measuring the DUT
Connect the DUT and perform measurements.
Settings of the electrical length and fixture compensation differ depending on the
measuring instrument used. Refer to the Operation Manual for the measuring instrument
that you are using.
NOTE
The
16196A/B/C/D requires frequent wear checks to keep the best measurement accuracy.
When the equipment is used for the first time following purchase or part replacement,
“Wear Check” should be conducted. Refer to “Wear Check” on page 44 for details.
Summary of Contents for 16196A
Page 7: ...6...
Page 18: ...17 2 Product Overview 2 Product Overview...
Page 37: ...36 Chapter3 Operation Connecting and Measuring DUTs...
Page 38: ...37 4 User Maintenance 4 User Maintenance...
Page 41: ...40 Chapter4 User Maintenance Overview Figure 4 1 Flowchart of User Maintenance...
Page 59: ...58 Chapter4 User Maintenance Assembling Check...