g. Record the sample probe resistance and bin number in the
Service Log at the end of the user’s guide; then press the STOP
button. The display will change to “Probe Bin Test”.
h. Press SETUP to enter the SETUP menu, then repeatedly press
the > button until the display reads “
Set Sample Bin #
”.
i. At “
Set Sample Bin #
”, press the START button. The display
will report the current sample bin number setting.
j. If the current sample probe bin setting is the same as that just
determined, simply press the STOP button twice to return to
“
Osmometer Ready
”. If the current sample bin number setting
is not the same as that just determined, enter the correct bin num-
ber by means of the numeric keypad, then press ENTER. The
display will change to “
Set Sample Bin #
”. Press STOP to return
to “
Osmometer Ready
”.
k. Press the CALIB button and calibrate the instrument as recom-
mended in Chapter 4. As advised in Chapter 4, the 3900 probably
will not allow automatic calibration at this time.
77
Troubleshooting & Service
Summary of Contents for 3900
Page 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Page 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Page 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...