The Advanced
®
Osmometer Model 3900 User’s Guide
80
“Block Probe
Failure”
“Block Probe
Open?”
“Calibration
Out of Range;
Repeat Calib”
“Calibrator Out
of Range”
“Cassette Index
Error”
“Cassette
Pusher Not
Back”
“Check
Calibration”
Switch the instrument off,
then on
Check the sample probe
by running the A/D Test.
Switch the instrument off,
then on.
Check the sample probe
by running the A/D Test.
Procedure, technique.
Wrong calibrator? See
Chapter 4.
Visually check for a cas-
sette in the wrong position
or other visible mechani-
cal cause.
Reported when “buzz”
point is changed. Buzz
setting can change cali-
bration.
Symptom
Check
Suggested Action
If the problem persists, see
Chapter 5 for further assis-
tance.
If the problem persists, see
Chapter 5 for further assis-
tance.
Repeat calibration.
Repeat calibration.
1. Clear the cause of the
problem, if visible, and
properly realign all
remaining cassettes
before continuing.
2. Press STOP to reset the
instrument.
3. Press START to continue
testing.
If the problem persists, see
Chapter 5 for further assis-
tance.
Recalibrate with new buzz
point.
Summary of Contents for 3900
Page 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Page 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Page 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Page 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...