or salt from your fingers can increase the electrical resistance of
the connection.
i. Arrange the leads so that they will not touch the clapper (see fig-
ure 7).
j. Align and adjust the probe and stir/freeze wire as instructed in
items 5 and 6 below, reset the probe bin number as instructed in
item 7, and recalibrate as instructed in Chapter 4.
5. Probe and Stir/Freeze Wire Alignment
Proper alignment and adjustment of the probe and stir/freeze wire
are of utmost importance to assure accuracy, precision and proper
operation of the instrument.
70
The Advanced
®
Osmometer Model 3900 User’s Guide
Figure 9:
Tube Gripper, Mandrel, Probe, Stir/Freeze Wire and
Alignment Tool End View
Tube
Gripper
Probe
Setscrew
Mandrel
Probe
Stir/Freeze
Wire
Alignment
Tool
Level B
Step