A.2 Auto-Calibration
Because errors in measurement and outputs will vary with time
and temperature, re-calibration is recommended when the module
is installed. Auto-calibration can measure and minimize errors
without external signal connections, reference voltages, or mea-
surement devices.
The PXIe-9834 has an on-board calibration reference to ensure
the accuracy of auto-calibration. The reference voltage is mea-
sured on the production line and recorded in the on-board
EEPROM.
Before initializing auto-calibration, it is recommended to warm up
the PXIe-9834 for at least 20 minutes and remove connected
cables.
Figure A-1: Auto-Calibration Block Diagram
NOTE:
NOTE:
It is not necessary to manually factor delay into applications, as
the PXIe-9834 driver automatically adds the compensation
time.
Analog Front End
ADC
Onboard Calibration Source
PXIe-9834
Switch
Signal Path /
Internal Calibration Source
ɃɃ
ɃɃ
ɃɃ
Summary of Contents for PXIe-9834
Page 6: ...vi Preface Leading EDGE COMPUTING This page intentionally left blank ...
Page 10: ...x List of Figures Leading EDGE COMPUTING This page intentionally left blank ...
Page 12: ...xii List of Tables Leading EDGE COMPUTING This page intentionally left blank ...
Page 17: ...Introduction 5 PXIe 9834 Figure 1 2 Typical Frequency Response 50Ω input impedance ...
Page 30: ...18 Introduction Leading EDGE COMPUTING This page intentionally left blank ...
Page 34: ...22 Getting Started Leading EDGE COMPUTING This page intentionally left blank ...
Page 60: ...48 Calibration Leading EDGE COMPUTING This page intentionally left blank ...
Page 64: ...52 Important Safety Instructions Leading EDGE COMPUTING This page intentionally left blank ...