2/5
B 320
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Connecting the detector directly to a printer
or another device
GB
00216
-
Edition
02
-
September
04
GB
00216
-
Edition
02
-
September
04
3/5
Alcatel Vacuum Technology France - ATH 200 User’s Manual
B 320
Connecting the detector directly to a printer
or another device
1
Configuration ticket
ASM1002 CONFIGURATION TICKET
VERSION: L125v1.0 r00
DATE:Jan/01/2003 TIME:00:00:07
SET-POINTS MENUS
audio level:
3
digital voice level:
3
hard vacuum alarm:
fixed
hard vacuum reject point:
1.0E-07
sniffing alarm:
fixed
sniffing reject point:
1.0E-04
sniffer probe clogged reject:
1.0E-06
bargraph zoom on reject point:
no
depollution function:
off
pollution reject point:
1.0E-05
antipollution GL function:
on
pollution reject point:
3.0E-04
memo. function active:
no
memo. display timer:
no
memo. timer value(min:s):
00:10
cycle end function:
operator
roughing overlap timer:
on
roughing timer value(min:s):
00:10
test timer value(min:s):
00:10
start.timer value(min:s):
00:05
background suppression
activation:
operator
trigger:
reject point
GL setpoint (mbar)
1.0E+02
Normal setpoint (mbar)
1.0E-01
SPECTRO MENUS
automatic calibration:
off
calibration check:
operator
every:
0050 cycles
every:
0010 hours
filament in use:
2
electronic zero:
110
acceleration voltage(V):
217
electronic current(mA):
0.0
sensitivity coefficient:
01.00
He calibrated leak
location:
internal
value:
1.0E-07
unit:
mbar.l/sec
year:
2003
loss per year(%):
2.00
temperature(C):
20
temp.coefficient(%/C):
3.00
MAINTENANCE MENUS
high vac.mnt.periodicity(hours):
12000
high vac. mnt.due in(hours):
12000
filament#1 running time(hours):
0
filament#2 running time(hours):
0
customed mnt.period.(cycles):
5.0E+05
customed mnt.due in(cycles):
5.0E+05
bicolore remote:
yes
primary pump1 used:
yes
primary pump2 used:
no
OTHER MENUS
test mode selection:
normal
inlet vent mode:
chamber
inlet vent delay(sec):
0
inlet vent open timer us:
no
inlet vent timer value(min/sec):
00:09
hard vacuum correction:
off
hard vacuum cor.coefficient:
1.00E+00
sniffing correction:
off
sniffing cor.coefficient:
1.00E+00
unit:
mbar.l/sec
display language:
english
user interface:
#4
password value:
5555
TYPICAL VACUUM VALUES
Pu_gf:1.00000
Mu_gf :1.0E+06
Pu_n :1.00000
Mu_n :00001.0
Mu_rld:00015.0
DATE AND TIME VALUES
last stop:
Jan/01/2003 00:00:00
last start:
Jan/01/2003 00:00:02
last calib.ok:
Jan/01/2003 00:00:00
detector counter (h:m:s):
00000:00:50