![background image](http://html.mh-extra.com/html/adixen/asm-182-td-plus/asm-182-td-plus_user-manual_2851543041.webp)
2/4
A 600
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Options
GB
02446
-
Edition
01
-
March
04
GB
02446
-
Edition
01
-
March
04
3/4
A 600
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Options
In order to reduce the roughing time when testing large volumes, a
second roughing pump can be added to the roughing system:
•
ASM 192 T / 192 T2 total capacity: 40 m
3
/h or 24 cfm.
•
ASM 192 TD+ / 192 T2D+ total capacity: 50 m
3
/h or 36 cfm.
Apart from the roughing capacity, the weight and the power
consumption, the characteristics and the use of the leak detector
remain the same.
Roughing system
7
The helium leak detector can be equipped with a software version
which will offer a complete RS 232 protocol:
•
3 operating modes: basic, advanced, printer;
•
possibility to remote control the detector (start/stop, autozero,
auto-cal etc...);
•
possibility to obtain and adjust the settings;
•
possibility to obtain all the maintenance information for preventive
maintenance purposes.
This RS 232 is the most effective interface to supervise your leak test
from a PC (data recording on an Excel sheet, for instance) and/or to
monitor the detector from a small PLC.
Interface board
8
Used to perform spray testing on long lines (typical diameter 1/4’’),
with a reduced response time due to the transfer of the helium by a
carrier gas injected in viscous flow.
In this case, the detector is equipped with an additionnal 1/4” VCR
connector specific to this option.
Test of gas line
10
3 lengths are proposed: 5 m, 10 m and 15 m.
Remote control cable
length
9
Designed for use of the unit in clean rooms (“Ultra Clean
Technology”).
The front and rear covers and frame are made of stainless steal.
Stainless steel cover
(UCT)
11
Alcatel Vacuum Technology France - User's Manual ASM 182/192 TD+ - ASM 192 T2D+