![background image](http://html.mh-extra.com/html/adixen/asm-182-td-plus/asm-182-td-plus_user-manual_2851543057.webp)
2/2
A 801
ASM 192 T2D+
Technical characteristics
GB
01365
- Edition
02
- Février
02
Start-up time (at 20 °C):
Without auto-calibration
3 min 30 s ± 10 %
With auto-calibration
5 min ± 10 %
Time to reach test mode (hard vacuum test):
Volume
46 l
200 l
1000 l
Configuration
Standard
2 ACP 28 Option
Standard
2 ACP 28
Standard
2 ACP 28 Option
Gross leak
43 s
23 s
2 mn 59 s
1 mn 32 s
6 mn 40 s
6 mn 21 s
Normal
1 mn 2 s
33 s
4 mn 25 s
2 mn 18 s
16 mn
9 mn 41 s
High sensitivity
1 mn 31 s
54 s
6 mn 30 s
3 mn 48 s
25 mn 35 s 16 mn 44 s
Residuel at 1.10
-8
1 mn 51 s
1 mn 16 s
8 mn 20 s
5 mn 42 s
31 mn 7 s
22 mn 7 s
Residuel at 2.10
-9
2 mn
1 mn 24 s
8 mn 56 s
6 mn 24 s
31 mn 1 s
24 mn 4 s
Miscellaneous:
Power voltage
low voltage:
high voltage:
100 - 130 V ± 10%
200 - 240 V ± 10%
Power frequency
50/60 Hz single phase
Power consumption*
1 ACP 28 Version 2 ACP 28 Version
maximum at roughing (delay 1 sec.):
1.5 kVA
2.3 kVA
stabilized at roughing (max.):
1 kVA
1.9 kVA
stabilized in test mode or in standby mode:
0.8 kVA
1.3 kVA
Start-up temperature
10 to 40° C
Ambient operating temperature
10 to 40° C
Storage temperature (standard detector)
-25° C to 70° C
Noise level (at 1 meter; audio alarm not operational, stand by mode)
66 dBA
House protection level
20 IP
Inlet
DN 50
Hygrometry max. (at 40°C)
95 %
Weight
1 ACP 28 version:
2 ACP 28 version:
157 kg
190 kg
Part to be tested weight max.
30 kg
*Average power consumption for a cycle on a 200 liters volume to obtain 1.10-8 mbar.l/s.
Alcatel Vacuum Technology France - User's Manual ASM 182/192 TD+ - ASM 192 T2D+