A 800
ASM 182 / 192 TD+
Technical characteristics
GB
01244
- Edition
03
- September
04
Displays and setpoints adjustments:
Inlet port pressure display range
10
3
to 10
-3
mbar / 10
5
to 10
-1
Pa
Audio alarm:
90 dB modulated and adjustable audio signal
Hard vacuum Audio signal set point
1.10
-19
to 9.9.10
19
Sniffing Audio set point
1.10
-19
to 9.9.10
19
Start-up time
(at 20 °C)
Without auto-calibration
< 4 min ± 10 %
With auto-calibration
< 5 min 30 s ± 10 %
Time to reach test mode
(Hard vacuum test):
inlet port
blanked-off
connected to
1.6 l volume
connected to
60 l volume
Gross leak test mode
2 s
< 6 s
< 1 min 20 s
High sensitivityl test mode
4 s
< 20 s
< 3 min 20 s
Miscellaneous:
Power frequency
low voltage:
100 - 130 V ± 10%
high voltage:
200 - 240 V ± 10%
Power frequency
50/60 Hz single phase
Power consumption (maximum)
1,5 kVA
Start-up temperature
10 to 40° C
Ambient operating temperature
10 to 40° C
Storage temperature (standard detector)
-25° C to 70° C
Storage temperature
(ASM 182 TD+ equiped
-20° C to 60° C
with the graphic interface control panel)
max. 48 h at -20° C
max. 168 h at 60° C
Humidity
(ASM 182 TD+
equiped with
the graphic interface control panel)
Ta
≤
40° C
Ta
>
40° C
85 % relative humidity max.
Absolute humidity must be
lower than the humidy of
85 % relative humidity at 40°
C (43,4 g/m
3
in the air)
Noise level (at 1 meter; audio alarm not operational, stand by mode)
65 dBA
House protection level
20 IP
Inlet
DN 40
Hygrometry max. (at 40 °C)
95 %
Weight
ASM 182 TD+ :
ASM 192 TD+ :
88 kg
195 kg
Alcatel Vacuum Technology France - User's Manual ASM 182/192 TD+ - ASM 192 T2D+