Specifications
28
Specifications
Stimulation Trigger
Main Interval
(1) START
A) MANUAL switch (front panel)
B) START INPUT (rear)
Input pulse: +3.5 V or over, 10
µ
s or over
Input impedance: 10 kilohms or over
C) Short-circuit of KEY terminal (rear) and E terminal
2) STOP
A) MANUAL switch (front panel)
B) STOP INPUT (rear)
Input pulse: +3.5 V or over, 10
µ
s or over
Input impedance: 10 kilohms and over
3) SET CYCLES
Setting range: 1 ms to 9990 s, within
±
0.1%
Unit: ms, s
Number: 3 digits
4) CYCLES
Repeat count: 1 to 999
Continuous stimulation: when set to “run”
5) External trigger
TRIG LEVEL: 0 to +10 V continuous variable
Approx. 3 V when turned fully counterclockwise
Input impedance: 100 kilohms or over
Stimulation Waveform Setup
1) DELAY
Setting range: 0 to 99.9 s,
±
1%, within
±
2
µ
s
Unit:
µ
s, ms, s
Number: 3 digits
2) INTERVAL
Setting range: 100
µ
to 999s, within
±
0.1%
Unit:
µ
s, ms, s
Number: 3 digits
3) DURATION
Setting range: 10
µ
to 99.9 s,
±
1%, within
±
2
µ
s
Unit:
µ
s, ms, s
Number: 3 digits