Specifications
32
Guidance and manufacture’s declaration - electromagnetic emissions
This Model 1001 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the 1001 should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-3
Complies
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The 1001 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The 1001 is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Guidance and manufacture’s declaration - electromagnetic immunity
This Model 1001 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the 1001 should assure that it is used in such an
environment.
Immunity test
Compliance level
IEC 60601 test level
Electromagnetic environment
-Guidance
Electrodatic discharge
(ESD)
IEC 61000-4-2
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user of the 1001 requires
continued operation during power mains
interruptions, it is recommended that the
1001 be powered from an
uninterruptible power supply or a
battery.
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
NOTE: Ut is the AC mains voltage prior to application of the test level.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
<5%
Ut
(>95% dip in
Ut
)
for 0.5 cycle
40%
Ut
(60% dip in
Ut
)
for 5 cycles
70%
Ut
(30% dip in
Ut
)
for 25 cycles
<5%
Ut
(>95% dip in
Ut
)
for 5s
<5%
Ut
(>95% dip in
Ut
)
for 0.5 cycle
40%
Ut
(60% dip in
Ut
)
for 5 cycles
70%
Ut
(30% dip in
Ut
)
for 25 cycles
<5%
Ut
(>95% dip in
Ut
)
for 5s
6 kV contact
8 kV air
6 kV contact
8 kV air
2 kV for power supply
lines
2 kV for power supply
lines
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode