20
Description of Parts
Name
4
Main start signal input terminal
(MAIN START)
5
Main key start terminal
(MAIN KEY START)
6
Stop signal input terminal
(STOP)
7
External trigger input terminal
(EXT TRIG)
Function
Input the signal to this terminal to commence main interval operation by an
external signal. It has the same function as the START button on the front panel.
The input signal must be a pulse waveform of +3.5 V or over with a width of 10
µ
sec or over.
When this terminal is short-circuited to the E (Earth) terminal, MAIN
INTERVAL operation commences.
The function is the same as the START button on the front panel.
Input the signal to this terminal to stop stimulation by an external signal. The
function is the same as the STOP button on the front panel. The input signal
conditions are the same as for START.
Input a signal to this terminal to trigger stimulation by an external signal.
Setting range: 0 to +10 V.
(See “External trigger level knob”)
/
Input conditions
Input impedance: approx. 100 kilohms
Allowable range of input voltage: 0 to +50 V
External Signal Input Terminals
[Note] Input the signals to the input terminals in the range specified in
the input conditions. If the signal exceeds the range, it may cause failure
or malfunction.
STOP
INPUT
MODU
MIX
MAIN
START
MAIN KEY
START
EXT
TRIG
EXT
TRIG
SPEAKER
OFF
ON
OUTPUT
INTERVAL
DELAY
MAIN
INTERVAL
SET CYCLES
END
TRAIN
END
7
6