Page 76 of 79
Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
ENGLISH
A.6 EMC COMPATIBILITY
Electromagnetic compatibility (EMC) is assessed with reference to the following standards:
IEC 60601-1-2:2007, 3rd edition
EMISSION
• EN 55011:2008+A2
• EN 61000-3-2:2006 +A1 +A2
• EN 61000-3-3:2008
IMMUNITY
• EN 61000-4-2:2009
• EN 61000-4-2/A1: 2010
• EN 61000-4-2
• EN 61000-4-3:2006
• EN 61000-4-4:2004
• EN 61000-4-5:2006
• EN 61000-4-5/A1
• EN 61000-4-6:2007
• EN 61000-4-8:2010
• EN 61000-4-11:2004
Guidance and manufacturer’s declaration – electromagnetic emissions
X-MIND unity is intended to be used in the electromagnetic environment specified below. The customer or the operator of X-MIND unity
must ensure that the device is used in this type of environment.
Emission test
Conformity
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
X-MIND unity uses RF energy only for inter-
nal operation. RF emissions are extremely
and attenuated are not likely to generate
interference with electronic equipment in
the vicinity.
RF emissions
CISPR 11
Class B
X-MIND unity is suitable for use in all esta-
blishments, including domestic establish-
ments and those directly connected to the
public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies