Page 63 of 79
ENGLISH
Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
Heating/Cooling Curves of the Toshiba D-041 anode
FIRMWARE
RXCD board
2,20
RXHVC board
2,14
PLEASE NOTE
Firmware versions maybe periodically updated therefore the versions indicated in the chart above may be different from the versions instal-
led on the unit.
To know the firmware versions installed in your X-MIND unity please follow the instructions in the following chapter “ADVANCED SETTINGS”
of the “X-MIND unity Installation and Maintenance manual”.
DEVICE POWER SUPPLY and ELECTRICAL CLASSIFICATION
Supply voltage
100 V à 240 V
Supply voltage frequency
50/60 Hz
Maximum Line Current
(technique factors that constitute maximum line current condition: 65kV, 7mA, 2s)
8,5 A (@100 V)
Standby current
100 mA
Maximum Adsorbed Power
850 VA
Maximum Rated Power
900 W
Adsorbed Power
(Stand-by mode, not armed)
3 VA
Adsorbed Power
(Armed mode – 240Vac/60Hz)
26 VA
Rated Power @ 0.1s
900 W à 70 kV/6 mA
Fuse (only Line)
T10 AH 250 V
Apparent Resistance
0.2 Ω