Page 57 of 79
ENGLISH
Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
E201
generator memory Inverter EEPROM error
NO
E202
generator memory Inverter I2C EEPROM error
NO
Code
Message
Description
Peut être réinitialisé
E203
generator memory Inverter EEPROM DMA readings fault
NO
E204
generator memory Inverter EEPROM DMA writings fault
NO
E205
generator not calibrated
Calibration not yet done
YES
E206
tube voltage too low
Anodic voltage too low
YES
E207
tube voltage too high
Anodic voltage too high
YES
E208
electrical discharge on hv
YES
E209
tube current too low
Anodic current too low
YES
E210
tube current too high
Anodic current too high
YES
E211
filament voltage too low
Filament voltage too low
YES
E212
filament voltage too high
Filament voltage too high
YES
E213
tube unit not connected
High voltage unit not connected
NO
E214
button released beforehand Enable signal released beforehand during an
exposure
YES
E215
generator internal power supply 12V internal voltage too low
NO
E216
generator internal power supply 12V internal voltage too high
NO
E217
generator mode
Wrong exposure enabling signal activation
YES
E218
safe circuit fail
Security interlock fault
NO
E219
tube current too low
Mean anodic current too low
YES
E220
tube current too high
Mean anodic current too high
YES
E221
no tube voltage feedback
Missing HV feedback
NO
E222
ACE acquisition fail
X-rays not detected by Sopix
YES
E223
tube unit temperature too high
Monoblock over temperature
YES
E224
tube unit temperature too low
Monoblock under temperature
YES
E225
tube unit temp. sensor fail
Monoblock temperature sensor fault (open
circuit)
YES
E226
tube unit temp. sensor fail
Monoblock temperature sensor fault (short
circuit)
YES
E227
voltage reference fail
generator voltage reference too low
YES
E228
voltage reference fail
generator voltage reference too high
YES
E232
system error
hardware error
NO
For the errors that can’t be cleared by the reset button on the keypad, please contact your installer or your local Acteon customer service